Device built-in
Comprehensive Analysis of PerkinElmer Atomic Absorption Spectrometer AAnalyst 800
1. Instrument Overview
The AAnalyst 800 from PerkinElmer is a high-performance atomic absorption spectrometer that integrates both flame (FAAS) and graphite furnace (GFAAS) modes, specifically designed for trace element analysis. Its core advantages include:
- Ultra-high sensitivity: Detection limits reach μg/L level in flame mode and ppt level in graphite furnace mode (e.g., cadmium detection limit of 0.008 μg/L).
- Fully automated operation: Supports one-click switching between flame and graphite furnace modes, with automatic optimization of optical path and sample introduction, significantly improving analysis efficiency.
- Modular expandability: Compatible with accessories such as hydride generators (HG) and flow injection systems (FIAS) to meet complex sample analysis needs.
This instrument is widely used in environmental monitoring, food safety, pharmaceuticals, and geological exploration, making it a benchmark device for trace element analysis in laboratories.
2. Core Technical Features
Optical System
- Dual-beam design: Simultaneous detection of sample and reference beams eliminates light source fluctuations (baseline stability ≤0.0006 Abs/30 min).
- Solid-state detector: Utilizes a full-spectrum, high-sensitivity CCD array with 70% quantum efficiency in the UV-visible range (189-900 nm), offering superior signal-to-noise ratio compared to traditional photomultiplier tubes.
Atomization System
- Flame mode: Corrosion-resistant all-titanium burner with direct slurry sampling capability for complex samples like milk powder and fruit juice.
- Graphite furnace mode:
- Transverse heating technology (THGA): Eliminates temperature gradients and reduces memory effects (e.g., molybdenum memory effect of only 3.3%).
- True Temperature Control (TTC): Real-time compensation for voltage and graphite tube resistance variations ensures stable heating rates (2000°C/s).
Background Correction Technology
- Longitudinal Zeeman effect: Flame mode uses deuterium lamp correction, while graphite furnace mode employs longitudinal alternating magnetic fields, increasing light intensity by 50% compared to transverse Zeeman.
Intelligent Software
- WinLab32 platform: Complies with 21 CFR Part 11 regulations, automatically generates reports, and adheres to GLP/GMP standards.
3. Typical Applications
- Environmental Monitoring
- Detection of heavy metals (e.g., lead, arsenic) in water (meets EPA standards) and soil pollution surveys (e.g., HJ 737-2015).
- Food Safety
- Trace analysis of cadmium in rice (GB 5009.15) and mercury in seafood, with detection limits as low as 0.1 μg/L.
- Pharmaceutical Industry
- Metal impurity testing in injectables (USP<232>), complying with ICH Q3D requirements.
- Industrial and Geological Applications
- Ore composition analysis (e.g., gold, copper) and impurity control in lithium battery materials.
4. Innovations and Industry Contributions
- High-Efficiency Design
- Graphite tube lifespan exceeds 3,000 tests, with argon-saving mode reducing operational costs.
- Regulatory Compliance
- ISO 9001 certified, supports CNAS/CMA laboratory accreditation.
- Hybrid Expansion
- Compatible with HPLC and GC for indirect analysis of sulfate, amino acids, etc.
5. Service and Support
PerkinElmer provides comprehensive support:
- After-sales response: Telephone support within 4 hours, core parts delivery within 7 days.
- Method development: Assistance in optimizing experimental conditions (e.g., soil testing case study at Jiaxing University).
Summary
The AAnalyst 800, with its high precision, automation, and versatility across multiple fields, continues to advance atomic absorption spectroscopy technology. Whether meeting stringent environmental monitoring requirements or complying with pharmaceutical industry standards, this instrument delivers reliable solutions, reinforcing PerkinElmer's leadership in analytical instrumentation.