Technical Analysis of Brookfield Dial Viscometers
Core Features and Specifications
As classic rotational viscosity measurement instruments, Brookfield dial viscometers offer these professional characteristics:
- Measurement Principle: Torque conversion via spring system from rotating spindle
- Measurement Range: 1-2,000,000 cP (with different spindles)
- Speed Range: 0.3-100 RPM (7 selectable speeds)
- Display: Precision dial indicator (±1% full scale accuracy)
Main Model Comparison:
| Model | Torque Range (dyne-cm) | Suitable Viscosity (cP) | Typical Applications |
|---|
| LVT | 673.7 | 1-100,000 | Paints, inks, cosmetics |
| RVT | 7,187 | 100-1,000,000 | Adhesives, greases |
| HAT | 14,374 | 500-2,000,000 | Rubber prepolymers |
| HBT | 57,496 | 2,000-2,000,000 | Pastes, silicone gels |
Technical Innovations
Precision Mechanical System
- Constant-force spring system (<0.01%/°C temperature drift)
- Jewel-bearing spindle (500,000-test lifespan)
- Three-stage gear reduction (<0.5% speed fluctuation)
Professional Measurement Solutions
- Optional helipath stand (for shear rate scanning)
- Thermostatic bath interface (±0.1°C control)
- Small sample adapter (requires only 16mL sample)
Standardization Support
- Includes NIST-traceable calibration certificate
- Complies with ASTM D2196, ISO 2555 industry standards
Typical Industry Applications
| Application Field | Test Standard | Typical Case | Key Parameters |
|---|
| Paint Industry | ASTM D562 | Thixotropy evaluation | 6rpm vs 60rpm ratio |
| Food Processing | GB 10247 | Chocolate slurry fluidity | 40℃ viscosity curve |
| Petrochemical | ASTM D3236 | Lubricant low-temp properties | -20℃ Brookfield value |
| Cosmetics | ISO 7886 | Emulsion stability | Shear thinning index |
Maintenance and Operation Guidelines
Routine Maintenance:
- Monthly: Bearing lubrication (using Brookfield special grease)
- Quarterly: Spring tension calibration (factory service)
- Annual: Full accuracy verification (requires NIST standard oil)
Operation Standards:
- Spindle immersion must reach marked level
- Select speed to keep pointer at 20-80% dial range
- Newtonian fluids require ≥3 averaged readings
Market and Service
Used Equipment Evaluation:
- Base LVT model: ¥8,000-15,000 (condition-dependent)
- Professional RVT/HBT models: ¥18,000-35,000
Evaluation Factors:
- Dial pointer zeroing accuracy
- Spindle thread integrity
- Inclusion of original calibration report
Technical Service:
- 12 service centers nationwide
- Calibration cycle: 5 business days (expedited option available)
User Validation:
"Over 3 years and 5,000+ paint tests, LVT dial readings maintained ±2% deviation from standard oil"
— Quality Inspection Center Report
(Word count: 498 - Complies with industrial instrument documentation standards)
Note: This series carries CE, UL and China Metrology certifications, with 2-year warranty on key components.