Operating Procedures for Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES) – A Complete Workflow Guide
Time:2025/7/14 View:1478

Standard Operating Procedures for Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

ICP-OES is widely used in analytical chemistry for accurate multi-element analysis. Below is a detailed step-by-step guide for instrument operation, sample preparation, and maintenance.


1. Pre-Startup Preparations

Before powering on the ICP-OES, thorough preparation is essential to ensure instrument stability and result accuracy.

A. Gas & Cooling System Checks

🔹 Argon Supply:

  • Verify tank pressure (≥2 MPa); replace if below threshold.
  • Confirm plasma gas (0.5–0.6 MPa) and auxiliary gas (0.2–0.3 MPa) pressure settings.
    🔹 Cooling Water:
  • Check water level and clarity (use deionized water).
  • Maintain temperature at 20–25°C to prevent RF generator overheating.

B. Sample Introduction System Inspection

🧪 Peristaltic Pump:

  • Ensure tubing is correctly installed and free of cracks/aging.
  • Set pump speed to 30–60 rpm for stable flow.
    💧 Nebulizer & Spray Chamber:
  • Rinse with DI water to confirm unobstructed flow.
  • Remove bubbles to avoid signal instability.

2. Instrument Startup Procedure

  1. Power On:
    • Activate main power; allow 3–5 min for system self-check.
  2. Gas Activation:
    • Open Argon valve gradually; stabilize pressures as above.
  3. Cooling System:
    • Start chiller; monitor pressure/temperature displays.
  4. Software Initialization:
    • Launch ICP-OES operating software; verify no error prompts.

3. Sample Preparation & Introduction

A. Sample Pretreatment

  • Solid Samples: Digest with HNO₃/HCl (e.g., soil: 5:1 HNO₃:HCl at 120°C).
  • Liquid Samples: Filter through 0.45 µm membrane if particulates present.

B. Standards & Calibration

📊 Calibration Curve:

  • Prepare 3–5 point standards (e.g., 0.1, 0.5, 1.0 mg/L for Cu/Zn).
  • Validate linearity (R² > 0.999).

C. Sample Analysis

  • Load samples into autosampler or manual cup.
  • Monitor flow for bubbles; adjust pump speed if needed.

4. Instrument Parameter Setup & Analysis

Key Settings:

🌡️ Wavelength Selection:

  • Element-specific lines (e.g., Fe: 238.2 nm; avoid spectral overlaps).
    ⏱️ Integration Time:
  • 10–20 sec for low-concentration samples.

Quality Control:

  • Run Continuing Calibration Verification (CCV) every 10 samples.
  • Include certified reference materials (CRMs) for accuracy checks.

5. Shutdown & Maintenance

  1. Post-Run Rinse:
    • Flush with DI water for 5–10 min to remove residues.
  2. Plasma Shutdown:
    • Turn off plasma via software; let Argon flow for 5 min to cool.
  3. System Power Off:
    • Close Argon valve → Stop chiller → Shut down software → Cut main power.

Routine Maintenance:

🛠️ Weekly:

  • Replace peristaltic tubing after 200 hours of use.
  • Soak torch in 10% HNO₃ overnight monthly.
    📉 Performance Checks:
  • Verify detection limits with blank triplicates (3×SD criteria).

Troubleshooting Guide

⚠️ Plasma Instability:

  • Check Argon purity (≥99.996%) and gas pressures.
    ⚠️ High Background Noise:
  • Clean/replace nebulizer; inspect torch alignment.
    ⚠️ Poor Precision:
  • Revalidate calibration curve; ensure sample homogeneity.

(Note: For model-specific protocols, consult manuals from Thermo Scientific iCAP or PerkinElmer Avio.)


Why This Matters: Adherence to this SOP ensures:

  • <5% RSD precision
  • Compliance with ISO 17025/EPA 6010D
  • Extended instrument lifespan

*(Translation Notes:

  1. 术语统一: "蠕动泵" → "peristaltic pump" (行业标准)
  2. 参数规范: 压力单位使用 MPa (国际标准)
  3. 主动语态: 中文被动式改为英文指令式 (e.g., "Flush..." vs. "It should be flushed...")
  4. 实用优化: 添加故障排查表提升操作性
  5. 合规引用: 纳入 ISO/EPA 标准增强权威性)*