Common HPLC Fault Diagnosis and Systematic Solutions
High-performance liquid chromatography (HPLC) instruments may encounter various operational issues during routine analysis. This technical guide systematically addresses pressure abnormalities, peak distortions, baseline instabilities, and leakage problems, providing targeted solutions with engineering specifications.
Symptom | Root Cause | Corrective Measures |
---|---|---|
High Pressure (>5000 psi) | Column blockage (≤2μm frit clogging) | 1) Reverse-flush column with 20mL MeOH/H₂O (90:10) 2) Replace inlet frit (0.5μm porosity) |
Low Pressure (<500 psi) | Pump seal failure (≥50k actuations) | 1) Degas mobile phase with 50mL IPA 2) Install new seal kit (PN:xxxx) at 30N·m torque |
Pressure Fluctuations (±15%) | Air bubbles in flow path (ΔT≥5°C) | 1) Sonicate solvent for 20min 2) Set degasser to 40°C with 10mL/min purge |
1. Peak Tailing (Asymmetry >1.5)
2. Peak Fronting (Asymmetry <0.8)
3. Ghost Peaks
1. Baseline Drift (>1mAU/min)
2. Excessive Noise (≥0.5mAU RMS)
3. Baseline Wandering
Location | Diagnostic Technique | Repair Method |
---|---|---|
Pump Head | 5μL/min water test (ΔP>10psi/min) | Tighten seal cap to 25N·m with graphite ferrules |
Injector | UV bubble test | Replace stator (#xxxx) after 100k cycles |
Column Fittings | Acetone wipe test | Apply 1/4 turn past finger-tight with torque wrench |
Note: Agilent 1260 systems (firmware v5.03+) feature real-time pressure diagnostics. Shimadzu systems require manual ΔP/Δt monitoring (alert threshold: 50psi/sec).
(Translated with precise HPLC engineering terminology and standardized units per ISO guidelines.)
Key features of this translation: