Technical Overview and Maintenance Guide for Waters SYNAPT G2-Si Mass Spectrometer
Technical Overview and Maintenance Guide for Waters SYNAPT G2-Si Mass Spectrometer
1. Instrument Overview
The Waters SYNAPT G2-Si is a high-performance quadrupole-time-of-flight (Q-TOF) tandem mass spectrometer featuring Triwave® ion mobility separation. It is widely used in proteomics, metabolomics, environmental contaminant analysis, and small-molecule structural elucidation in complex matrices. Its core advantages include:
- Multidimensional separation: Mass-to-charge (m/z), ion mobility (CCS), and MS/MS fragments
- High resolution: >40,000 FWHM
- Ultrahigh sensitivity: Sub-femtogram detection limits
2. Core Technologies & Configurations
2.1 Key Modules
- Ion Sources: Supports ESI (electrospray ionization) and APCI (atmospheric pressure chemical ionization) for diverse sample types.
- Triwave® Ion Mobility: Separates ions based on collision cross-section (CCS), enhancing peak capacity.
- StepWave™ Ion Optics: Maximizes ion transmission while reducing noise (S/N improvement >10x).
- Tof Reflector: Delivers <2 ppm mass accuracy for precise compound identification.
2.2 Software Platform
- UNIFI® Scientific Information System: Integrates data acquisition, processing, and database searching (targeted/non-targeted workflows).
- Dynamic Range Enhancement (DRE): Enables **>5 orders of magnitude** dynamic range.
3. Routine Maintenance Guide
3.1 Daily Maintenance
- Ion Source Cleaning:
- Disassemble ESI/APCI probe, sonicate in isopropanol (10 min), rinse with HPLC-grade water, and dry with N₂.
- Check capillary for clogging; flush with MeOH/H₂O (1:1) if needed.
- Vacuum System Check:
- Confirm mechanical pump oil level (replace if cloudy).
- Verify turbopump speed (>90% rated) and TOF pressure (<1×10⁻⁶ mBar).
3.2 Weekly Tasks
- Mass Calibration:
- Use NaCsI or leucine enkephalin calibrant; ensure <2 ppm mass error via UNIFI.
- Gas System Inspection:
- Maintain N₂ pressure (7–10 psi for ESI); replace desiccant if color-indicator turns pink.
3.3 Monthly Maintenance
- Ion Lens Cleaning:
- Soak and sonicate ion guides in MeOH/H₂O (1:1); avoid scratching surfaces.
- Pump Oil Replacement:
- Change every 500 hours (use Edwards E2M28 oil for rotary vane pumps).
3.4 Annual Professional Service
- Schedule turbopump maintenance, high-vacuum seal replacement, and full-system performance validation with Waters engineers.
4. Troubleshooting Common Issues
Symptom | Possible Cause | Solution |
---|
Sensitivity drop | Ion source contamination | Clean ESI probe; optimize lens voltages |
Mass accuracy drift | Temperature fluctuation | Recalibrate with fresh standard |
Vacuum instability | Pump oil degradation | Replace oil; check for leaks |
5. Critical Best Practices
- Avoid frequent power cycles: Allow **>4 hours** for turbopump cooldown to prevent bearing damage.
- Use LC-MS grade solvents only: Minimize salt/particulate contamination.
- Back up method files: Export instrument parameters via UNIFI to external storage.
6. Conclusion
Proactive maintenance of the SYNAPT G2-Si—particularly for the ion source and vacuum system—ensures long-term reliability and data integrity. Laboratories should:
✓ Maintain a detailed service log
✓ Follow Waters-recommended intervals
✓ Engage technical support for complex issues
Appendix:
- Recommended consumables: Waters P/N 700004690 (ESI probe), 700003208 (capillary)
- Emergency support: Waters Technical Service Hotline (400-820-2676 in China)
Note: Adjust maintenance frequency based on usage intensity and lab environment.
Key Performance Metrics:
▶ 99.7% uptime with scheduled maintenance
▶ <1% CV in retention time reproducibility
▶ 5+ years typical service life under optimal care
(Compliant with GMP and 21 CFR Part 11 requirements)
Why This Matters:
- Pharma Labs: Meet FDA data integrity standards
- Academic Users: Protect years of calibration data
- CROs: Minimize downtime in high-throughput workflows